Scientific Publications Database

Article Title: Small composite field correction factors for the CyberKnife radiosurgery system: clinical and PCSR plans
Authors: Christiansen, E.; Muir, B.; Belec, J.; Vandervoort, E.
Journal: PHYSICS IN MEDICINE AND BIOLOGY Volume 62 Issue 24
Date of Publication:2017
Abstract:
A formalism has been proposed for small and non-standard photon fields in which k(Qclin, Qmsr) (fclin, fmsr) correction factors are used to correct dosimeter response in small fields (indiviual or composite) relative to that in a larger machine-specific reference (MSR) field. For clinical plans consisting of several fields, a plan-class specific reference (PCSR) plan can also be defined, serving as an intermediate calibration field between the MSR and clinical plans within a certain plan-class. In this work, the formalism was applied in the calculation of k(Qclin, Qmsr) (fclin, fmsr) for 21 clinical plans delivered by the CyberKnife radiosurgery system, each plan employing one or two of the smallest diameter collimators: 5 mm, 7.5 mm, and 10 mm. Three detectors were considered: the Exradin A16 and A26 micro chambers, and the W1 plastic scintillator. The clinical plans were grouped into 7 plan-classes according to commonly shared characteristics. The suitability of using a PCSR plan to represent the detector response of each plan within the plan-class was investigated. Total and intermediate correction factors were calculated using the egs_chamber Monte Carlo user code. The corrections for the micro chambers were large, primarily due to the presence of the low-density air cavity and the volume averaging effect. The correction for the scintillator was found to be close to unity for most plans, indicating that this detector may be used to measure small clinical plan correction factors in any plan except for those using the 5 mm collimator. The PCSR plan was shown to be applicable to plan-classes comprising isocentric plans only, with plan-classes divided according to collimator size. For non-isocentric plans, the variation of k(Qclin, Qmsr) (fclin, fmsr) as a function of the point of measurement within a single plan, as well as the high inter-plan-class variability of the correction factor, precludes the use of a PCSR plan.